About Tedlar Bag

Blank Test

 

A Tedlar Bag (1L) was filled with nitrogen gas and put in an oven at 60 °C. 1 hour later, the gas in the bag was withdrawn into a gastight micro syringe and injected directly to a GC/MS. The bag was not flushed / cleaned in advance. 

 

Aldehydes/Acetone Blank Test

 

After being filled with nitrogen gas, a Tedlar bag was closed and put in an oven at 60 °C. 1 hour later, 1L gas in the bag was vacuumed with a pump through a DNPH cartridge and eluted sample was analyzed by HPLC. These bag was not flushed / cleaned in advance. 

 
 

Aldehydes/Acetone Blank Test

 

Vaporized standards as described below were introduced into Tedlar Bags. Then nitrogen gas was added into the Tedlar Bags and were measured by a GC/FID.

Standard Compound List: Acetone, Methylethylketone, Methyl isobutyl ketone, Isopropyl alcohol, Isobutyl alcohol, 1-Butanol, Ethylbenzene, Cyclohexane, n-Hexane, o-Xylene, m-Xylene, Toluene, Dicholoromethane, Methyl acetate, Ethyl acetate, Isobutyl acetate, n-Butyl acetate (50ng/mL each)

 
 
 
 

Storage Stability of Malodorous Compounds in Tedlar Bag

 

Each concentration of malodorous compound was prepared as described below and filled in a 5L Tedlar Bag together with nitrogen gas. Concentration changes over time were calculated with a detecting tube at certain times.

 

Gas Permeability Test

 

Permeation rate of O 2 , N 2 , CH 4 and CO 2 was measured using a permeation rate measurement system on Tedlar Bag. (in cm 3 /(m 2  • 24 h • 1 atm)


Material Hydrogen Oxygen Nitrogen Carbon dioxide
Tedlar Bag 324.9 20.4 2.9 158.5
 